EasyManua.ls Logo

Sonoscape S2 - M Mode Scanning

Sonoscape S2
8 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
- Volume: Press button, choose Volume (Height x Width x Length).
- Angle: Press button, choose Angle.
9. How to get IMT and specialized measurement
Press button to enter into the Calculation List, choose IMT under VASCULAR measurement.
Other specialized measurement can also be found under this list.
Note:
a. Both basic and specialized measurements can be done not only after image frozen but also
under real-time.
b. The calculation list is fully opened which offers measurements in different mode; for example,
you can get access to OB measurement even when youre in the vascular mode with linear
probe.
c. Trackball and key are used for choosing different calculation list or measurements.
M MODE SCANNING
10. How to change the B/M format?
In the 2D scanning, press to enter B/M mode. Press (MENU), use (Audio)
knob to choose items in the menu list. Change the B/M display format with menu knob
including five options. V1/1, V1/2, H2/1, H1/1, H1/2.
11. How to make the M measurement?
Time: press to enter the basic measurement, place trackball on the M graph, and then
move to the measurement list and click time.
Acceleration: press to enter the basic measurement, place trackball on the M graph, and
then move to the measurement list and click Slope.
HR: press to enter the basic measurement, place trackball on the M graph, and then move
to the measurement list and click HR.
Items under Basic measurement Time, Distance, Slope, T1/T2, D1/D2
S2 Operation Manual

Other manuals for Sonoscape S2

Related product manuals