SORVALL®
Centrifuges
5-33.
Access
Panel Removal
1.
Unplug
the
centrifuge
power cord.
2.
Remove
the
four screws
that
secure
the
access panel
to
the left
side
of the
centrifuge.
NOTE
The
lower
screw
towards
the
rear
of
the
centrifuge
has
a
grounding
lockwasher
which
must
be
replaced
when reinstalling
the
access panel.
3.
Lift
the
edge
of the
panel
until
it
unlocks
from the
top
plate
and
remove
the
panel.
4.
To
replace
the
access
panel,
reverse
steps
2
through
3.