9-11
XPREZZON AND QUBE
O
PERATIONS MANUAL
TRENDS
To set the scale for a measurement
1 From the Trends Setup window, touch the tab for the graph that you want to
adjust the properties for.
2 From the measurement list, touch the measurement that you want to set the
scale for.
3 In the Scales box, touch the Min or Max text entry boxes.
A keypad shows where you enter the scale value.
To add reference lines for a measurement
1 From the Trends Setup window, touch the tab for the graph that you want to
adjust the properties for.
2 From the measurement list, touch the measurement that you want to add
reference lines to.
3 In the Reference Lines box, touch the Min or Max check boxes to enable the
minimum and maximum reference lines for the measurement.
4 In the Reference Lines box, touch the Min or Max text entry boxes to set the
position of the reference lines for the measurement.
A keypad shows where you enter the value of the Min or Max.