2015 Spectral Evolution, Inc. 1 Canal St. Unit B-1 Lawrence, MA 01840 USA
13
PSR Series Spectroradiometer Manual
Revision: 1.08
3/31/2015
To erase all memory and set MEM# to one (1), the <MENU> and <EDIT> keys must be
pressed simultaneously. IMPORTANT! This operation permanently erases all internally
stored scans taken in standalone operation mode.
STIM xxx - STIM xxx is the scan averaging time. The allowable values are from 1 to 30.
The length of time for a scan is xxx seconds, for each light and dark scan.
O:xxxxxx - Entrance optic option. This must be set to the entrance optic in use for correct
calibration factors to be applied. There can be up to eight (8) optic calibration options
stored in a single instrument. Some possibilities are listed below:
(1) LENS 4 4 degree FOV lens
(2) FIBER1 standard 25 degree FOV fiber optic
(3) LENS14 14 degree FOV lens
(4) RAW DN no calibration applied
LSCN xxx - Laser Scan Switch function. This can be set to ON or OFF. When the Laser
Switch is depressed, the sighting laser is (always) turned on. When LSCN is set to ON,
a stand alone scan will be acquired when the switch is released. When LSCN is set to
OFF, no scan will be acquired.
3.3.4 Modifying a parameter
(1) Depress the MENU button to locate the parameter you wish to edit.
(2) Modify the parameter by depressing the EDIT button. The push button switches are
programmed to scroll the menu of available options.
(3) Depress the MENU button again to save the selected parameter setting.