While in ZERO mode, press HOLD to freeze the
reading. The open hand icon will appear.
While in HOLD press PEAK to display the highest
readings recorded during the ZERO mode. The
mountain peak icon will appear.
Press LIVE to return to ZERO mode, and then press
NORM to return to NORMAL mode.
8.4 INTEGRATION (INTG) MODE
In INTEGRATION mode, the meter sums up the
cumulative ultraviolet light that the sensor is exposed
to over an interval of time, and displays the resultant
total energy in joules.
While in NORMAL mode, press INTG. All values will
be set to zero, the hourglass icon will appear. Top
line display includes UV wavelength and real-time
intensity reading (autoranging between µW/cm
2
and
W/cm
2
).
Press START to begin real-time INTEGRATION.
Timer starts counting in seconds as indicated by
running man icon.
While integrating, press FREEZE to freeze and hold
the displayed data. The snowflake icon will indicate
that the display is static.
The AccuMAX will continue the INTEGRATION
process in the background, as indicated by the
circular arrows icon.
Press RESUME to display the ongoing integration
readings.
Press STOP to end INTEGRATION cycle.
Press NORM to return to NORMAL mode.
NOTE: While integrating, the AccuMAX will not shut
off automatically after 60 minutes as it would in other
modes and settings. This allows the user to continue
integrating for extended periods of time.