Display Screens
Initial Power Up
When the monitor is first powered up, all front panel lights turn ON and a splash screen appears showing
the monitor’s current firmware revision level. After a brief moment the Warm Up screen is display along
with the front panel ON light (green) blinking.
The monitor requires 900 seconds (15 minutes) to warm up (USER MAY VIEW LCD DISPLAY COUNT-
DOWN FRO
M 900 SECONDS); after which, the ON light glows steady and the Data Display screen is
displayed.
Filling the Purge-Air Bag
WARNING: Fill the purge-air bag in a clean, fresh-air environment that is at the same
temperature and relative humidity as the area being check for gas. DO NOT fill the bag in an
area that is contaminated with sulfuryl fluoride gas.
The following message is displayed when the monitor’s purge air bag is empty and needs refilling. Press the
ENTER button to start the refill process. After the air bag has been filled, the monitor will automatically
resume normal operation.
SPECTROS
VERSION 2.14
WARM UP
900
PURGE AIR EMPTY
<ENTER> TO FILL
RECHARGING PURGE
AIR...
9