IM-P505-02-US Issue 1
76
UTM20 Series Ultrasonic Transit-time Flowmeters
Out-of-Specification Codes
Code Description Correction
S01 ELECTRONIC
WARNING
Fault detected and meter rebooted
Contact factory, update firmware, or
repair or replace transmitter
S02 DEFAULT FAILED Reset to factory defaults failed
Check calibration If it does not
match the calibration settings on
the transducer serial tag, enter
field calibration settings Return
to the Home Screen and continue
to operate (if the reset to factory
defaults is through the transmitter)
S03 LANGUAGE FILE
CORRUPT
English only Update firmware
S10 mA TOO HIGH
Flow or energy rate higher than flow
rate at 20 mA output
Check the scaling of the Current
#1 output
S11 mA TOO HIGH
Flow or energy rate higher than
20 mA
Check the scaling of the Current
#2 output
S19 mA SUPPLY VOLTAGE
ERR
Supply voltage out of range for 4-20
mA outputs
Check wiring
S20 FREQ HIGH
Value higher than max frequency
output
Check the scaling of the frequency
on Output #1
S21 FREQ HIGH
Value higher than max frequency
output
Check the scaling of the frequency
on Output #2
S22 FREQ HIGH
Value higher than max frequency
output
Check the scaling of the frequency
on Output #3
S30 PULSE HIGH
Pulse output is too fast for the pulse
width
Check the scaling factor, units and
pulse width of the pulse on Output
#1
S31 PULSE HIGH
Pulse output is too fast for the pulse
width
Check the scaling factor, units and
pulse width of the pulse on Output
#2
S32 PULSE HIGH
Pulse output is too fast for the pulse
width
Check the scaling factor, units and
pulse width of the pulse on Output 3
S33 PULSE HIGH
Pulse output is too fast for the pulse
width
Check the scaling factor, units and
pulse width of the pulse on Aux
Output #1 dry contact
S34 PULSE HIGH
Pulse output is too fast for the pulse
width
Check the scaling factor, units and
pulse width of the pulse on Aux
Output #2 dry contact
S40 HIGH FLOW
Flow rate is above high flow alarm
setting
Check flow rate and Set High setting
for Output #1
S41 HIGH FLOW
Flow rate is above high flow alarm
setting
Check flow rate and Set High setting
for Output #2
S42 HIGH FLOW
Flow rate is above high flow alarm
setting
Check flow rate and Set High setting
for Output #3