EasyManua.ls Logo

Spirax Sarco UTM20 Series - Transmitter Specifications; Inputs, Outputs, and Ports Specifications; Data Logging, Alarms, Languages, and Security

Spirax Sarco UTM20 Series
108 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
IM-P505-02-US Issue 1
87
UTM20 Series Ultrasonic Transit-time Flowmeters
Flow Meter Energy Meter
Inputs and
Outputs
0/4 to 20 mA
output
One 16-bit, isolated, max 800
Ohms, internal or external power
Two 16-bit, isolated, max 800
Ohms, internal or external power
Digital input
One 5 to 30V DC, isolated, externally or internally sourced, reset
totalizer or alarm output
Digital output
Two selectable pulse, alarm,
flow direction, sink isolated open
collector, 5 to 30V DC, max
50 mA externally or internally
sourced, leakage current 1uA
max
Three selectable pulse,
frequency, alarm, flow direction,
isolated open collector, 5 to
30V DC, externally or internally
sourced, leakage current 1uA
max
Frequency output: 50% duty cycle, 63 to 10k Hz maximum frequency
Pulse (totalizer) output: 5 kHz max output, open collector, pulse width
5 to 500 ms programmable
Optional: Two dry contact output for alarm or flow direction 30V DC
max , 5A max (Ethernet not available with this option)
RTD (energy
only)
None
Two 2-wire, 3-wire or 4-wire
Pt100/Pt1000 RTD 12-bit inputs;
Range of -40 to 200 °C; Clamp-
on resistor kits available
Ports
Programming
USB 2 0 mini B connector for connection to a device with SoloCUE
configuration software
EIA-485
Modbus RTU command set or BACnet MS/TP; Baud rates 9600,
14400,19200, 38400, 57600, 76800, 115k; terminating resistor
selectable
Ethernet
Optional 10/100 Base T RJ45, communication via Modbus TCP/IP or
BACnet/IP
AquaCUE/
BEACON
Connectivity to AquaCUE/BEACON endpoint (LTE cellular)
Data Logging
Number of points
Up to 8 parameters per record Selectable 1 second to 1 day Transfer
logs via memory card
Real Time Clock
Backed up with a super capacitor, minimum of 32 days of data
retention without power; Requires no servicing
MicroSD card
slot
8 GB card, included with transmitter
Alarms Records 150 previous alarms, warnings or errors
Languages English, French, German, Italian, Spanish
Security
Four levels: Read-only, Operator, Service and Admin; 6-digit passcode number; selectable
auto logout