12 Stat Fax 303 Plus Operator’s Manual Rev. J
Awareness Technology, Inc.
1.8 Specifications
Photometric
Linear Measurement Range ......................0.00 to 3.0 Absorbance Units (A)
Photometric Accuracy ................................± (1% of the reading + 0.01 A) from 0 to 1.5A
................................................................... ± (2% of the reading + 0.01 A) from 1.5 to 3.0A
Stability ......................................................Drift of no more than 0.005A in 8 hours
Warm up Time ...........................................45 seconds, built-in
Light Source...............................................Tungsten Lamp
Standard Wavelengths...............................Standard: 405, 450, 492, 630nm
Six Filter UV: 405, 450, 492, 630, 545, and 340 nm
Six Filter VIS: 405, 450, 492, 630, 600, and 545 nm
Type of Filter ..............................................3-cavity hard coat interference filter, 10nm half band
pass typical
Vessel ........................................................Single, double, or break apart strips, up to 12 wells
long,
3 strip loading capacity
Electronic and Software
Speed ........................................................Reads, calculates and prints results for up to 12 wells
in about 30 seconds
Display .......................................................LCD, 16” characters, 5x7 dot matrix
Internal Printer ...........................................Thermal, dot matrix, 20 characters per line
Keyboard ...................................................16-key, membrane switch, 4X4
Calculation Modes .....................................Single point calibration, Multi-point calibration with
regression or point-to-point curve fit, cutoff, cutoff
control, % Absorbance Multi-Point, Uptake.
Serial Port ..................................................Output only, 2400 Baud, 1 start bit, 8 data, 1 stop, no
parity, no handshake
Power Source ............................................Switch selectable power supply (115V or 230V
indicated)
Voltage Source 110-120/220-240 VAC from 50 to 60
Hz, CAT II
Power consumption is less than 50 Watts
Fuse: two 1/2A, T rating, 250V, 3AG type fuse
All power cords must be approved for the country of
use
Certifications and Compliances .............NRTL Listed Standard UL -61010A-1, Electrical
Equipment for Laboratory Use.
CE Certificate of Conformity. Conforms to the
following standards: EN 61010-1, EN 61326, following
the provisions of the 98/79/EC IVD directive
Physical
Enclosure...................................................ABS Plastic cover with metal base
Dimensions .............................................Approx. 9”x 12” x 3”, (23 x 30 x 8 cm) 13 lbs (5.9 Kg)