OPERATOR MANUAL
Reliance
®
400 Laboratory Glassware Washer
Reliance
®
500 Laboratory Glassware Washer
(2008-03-31) P122994-513

Table of Contents

Questions and Answers

  • J
    Jerome JordanSep 23, 2025
    What to do if the Steris Reliance 500 shows empty chemical container?
    • A
      Andrew ClementsSep 23, 2025
      Press ALARM REPLY touch pad to silence alarm buzzer. Install new supply and press CYCLE/START touch pad to resume operation.
  • S
    Seth MurphyAug 23, 2025
    How to fix insufficient water entering the Steris Reliance 500 chamber during treatment?
    • R
      Robert RobertsAug 23, 2025
      If there is insufficient or no water entering the chamber during treatment, make sure that the supply valves are fully open. The supply line pressure must meet the specifications outlined in the equipment drawing. Open the building and washer supply valves.
  • C
    Christopher RandolphAug 25, 2025
    What to do if the Steris Reliance 500 Washer chamber does not drain completely?
    • S
      sullivanedwardAug 25, 2025
      If the chamber isn't draining completely, start by checking the chamber debris screens for clogs and clean them if necessary. Also, flush out the drain line to ensure it is clear. Verify the building piping and flush it out if necessary.
  • J
    Joshua DuncanAug 27, 2025
    Why is there foam in the Steris Washer chamber?
    • A
      Angela HendricksAug 27, 2025
      If you observe foam in the chamber, it might be caused by using the wrong type of detergent. Refer to the detergent specifications and follow the detergent manufacturer's recommendations for the amount of detergent to be used. Additionally, too much chemical injected during treatment can cause foam. Verify the chemical injection time in Automatic mode.
  • S
    Stephanie CunninghamAug 27, 2025
    What to do if the Steris Washer cycle does not start?
    • S
      Sandra NelsonAug 27, 2025
      If the cycle does not start when you press the CYCLE/START touch pad, or if the pump doesn't start during treatment, ensure that the chamber door is closed. If the door was open, close it securely and then press the CYCLE/START touch pad again.
  • K
    kaiserangieSep 17, 2025
    How to fix Steris Washer interior light that does not turn on?
    • N
      Nathan GrahamSep 17, 2025
      If the interior light does not turn on, first check if the POWER-OFF/STANDBY switch is in the OFF/STANDBY position and switch to POWER position. If the unit is still in Service mode, switch the washer to Automatic mode by pressing the EXTEND CYCLE touch pad. If the bulb is burned out, replace it.
  • C
    christinaponceSep 14, 2025
    How to fix Steris Reliance 500 water temperature in chamber sump is below programmed temperature?
    • D
      Dylan PriceSep 14, 2025
      If the water temperature in the chamber sump is below the programmed temperature, open the building and washer steam supply valves and press the CYCLE/START touch pad to resume operation. Also, verify steam traps in facility supply lines, if there is excess condensate in building supply lines.
  • M
    maymonicaSep 15, 2025
    How to troubleshoot if Steris Washer is equipped with double door, unload door fails to open?
    • K
      Kelsey GoodmanSep 15, 2025
      If the unload door fails to open on a washer equipped with a double door, it's likely due to the safety interlock feature being engaged. The unload door will not open until the washer has run one complete cycle.
  • H
    Hector WestSep 6, 2025
    What to do if the Steris Reliance 500 has too much water entering the chamber?
    • J
      James WoodSep 6, 2025
      If too much water is entering the chamber during treatment, causing it to go to the overflow drain, verify that the water level sensor float moves freely and clean it if necessary.
  • J
    joyceburnsSep 3, 2025
    How to fix a Steris Reliance 500 Washer chamber sump that is taking too long to drain?
    • B
      Brittany PruittSep 3, 2025
      If the chamber sump is taking too long to drain, verify the water level sensor and clean the sensor if necessary. Also, open the building and washer air supply valves and press CYCLE/START touch pad to resume operation if the air pressure is below specification.

Summary

1 Safety Precautions

Key Safety Warnings

Covers personal injury, equipment damage, electric shock, and burn hazards with specific warnings.

Additional Safety Warnings

Fire, Chemical, Burn, and Slipping Hazards

Details warnings related to fire, chemical exposure, burns, and slipping hazards.

Equipment Damage Cautions

2 Installation Verification

Technical Specifications

Details technical specs like voltage, amperage, power consumption, noise, and environmental conditions.

Installation Checklist and Environmental Factors

Installation Checklist Items

Provides a checklist for ensuring correct and complete installation.

3 Component Identification

4 Operating Instructions

Before Operating Equipment

Steps to take before operating the glassware washer, including safety checks.

Priming Procedure

Loading Glassware Washer

Typical Cycle Operation

Acknowledge Alarm Condition

Power Door Operation

5 Cycle and Control Value Programming

General Programming Overview

Introduction to cycles, treatments, and programmable values.

Change Values Mode Overview

Adding Cycle Treatments

Creating Custom Cycles

Removing Treatments From Cycles

Changing Miscellaneous Values

Selecting Injection Time

Verifying Chemical Concentration

Programming With Access Code

6 Routine Maintenance

Preventive Maintenance Schedule

7 Troubleshooting

No Alarm Screen or Printout Issues

Alarm Screen or Printout Issues

Steris Reliance 500 Specifications

General IconGeneral
BrandSteris
ModelReliance 500
CategoryWasher
LanguageEnglish