blairdawnAug 31, 2025
What to do if the Steris Washer treatment stops for no apparent reason?
- JJulia HartAug 31, 2025
If the treatment stops unexpectedly, check that the chamber doors are fully closed and close them securely.
What to do if the Steris Washer treatment stops for no apparent reason?
If the treatment stops unexpectedly, check that the chamber doors are fully closed and close them securely.
How to resolve insufficient water entering Steris Washer chamber through spray arms?
To resolve insufficient water entering the chamber through the spray arms, first ensure each accessory header is positioned directly over a manifold connector. Additionally, check the rotary spray arms and chamber debris screens for clogs and clean them if necessary.
How to resolve water leaks from the Steris Washer unit?
To address water leaks from the unit, tighten all clamps and connections.
What to do if the Steris Washer cycle does not start?
If the cycle does not start when you press the CYCLE/START touch pad, or if the pump doesn't start during treatment, ensure that the chamber door is closed. If the door was open, close it securely and then press the CYCLE/START touch pad again.
Why is there foam in the Steris Washer chamber?
If you observe foam in the chamber, it might be caused by using the wrong type of detergent. Refer to the detergent specifications and follow the detergent manufacturer's recommendations for the amount of detergent to be used. Additionally, too much chemical injected during treatment can cause foam. Verify the chemical injection time in Automatic mode.
What to do if the Steris Reliance 500 Washer chamber does not drain completely?
If the chamber isn't draining completely, start by checking the chamber debris screens for clogs and clean them if necessary. Also, flush out the drain line to ensure it is clear. Verify the building piping and flush it out if necessary.
What to do if the Steris Reliance 500 loads come out dirty?
If the loads are coming out dirty, it could be due to several reasons. First, check if the chemical supply is empty and install a new supply if needed. Ensure that you are using the correct type of detergent by referring to the detergent specifications and following the manufacturer's recommendations for the amount to be used. Also, inspect the rotary spray arms for clogs and clean them if necessary.
How to fix Steris Reliance 500 water temperature in chamber sump is below programmed temperature?
Covers personal injury, equipment damage, electric shock, and burn hazards with specific warnings.
Details warnings related to fire, chemical exposure, burns, and slipping hazards.
Details technical specs like voltage, amperage, power consumption, noise, and environmental conditions.
Provides a checklist for ensuring correct and complete installation.
Steps to take before operating the glassware washer, including safety checks.
Introduction to cycles, treatments, and programmable values.