- 21 -
increased, the stimulation level must be set to avoid the
muscular contraction.
Mode Constant C – Both Pulse Rate and Pulse Width
are adjustable. Default values are PR=70Hz and
PW=50μs characterized by an endorphins stimulating
effect through the Gate-control mechanism. Use this
program in case of ACUTE PAIN. The stimulation level
must be set to avoid the muscular contraction.
Mode MODulation SW (A) – Pulse Rate SWeep
modulation. Pulse Rate is automatically modulated within
F1=50Hz and F2=80Hz (adjustable); each frequency step
will last 1 sec. F2 can be changed only if the intensity
level of both channels are set to 0. The Pulse Width is
200μs (adjustable from 50μS to 250μS).
Mode MODulation 2F (B) – Pulse Rate Dual
Frequencies modulation. Pulse Rate is automatically
changed every 5 seconds, between F1=70Hz (adjustable)
and F2=5Hz. F1 can be changed only if the intensity level
of both channels are set to 0. The Pulse Width is 50μs
(adjustable from 50μS to 250μS).
Mode Burst A – 3 second Burst sequence = 1s on, 2s off.
Both Pulse Rate and Pulse Width are adjustable. Default
values are PR=30Hz and PW=250μs. The stimulation
level must be set to produce a gentle muscular
contraction.
Mode Burst B – 1.5 second Burst sequence = 0.5s on,
1s off. Both Pulse Rate and Pulse Width are adjustable.
Default values are PR=50Hz and PW=200μs. The
stimulation level must be set to produce a gentle
muscular contraction.