Electromagnetic compatibility
Instructions for use • OR1 SmartScreen • PQJ219_EN_V1.0_04-2021_IFU_CE-MDR 18
8 Electromagnetic compatibility
8.1 General notes on the operating environment
The product is suitable for use in professional healthcare settings. Professional healthcare
facilities include physician offices, dental offices, limited care facilities, freestanding surgical
centers, freestanding birth centers, multiple treatment facilities, hospitals (emergency rooms,
patient rooms, intensive care, surgical rooms, outside the HF-shielded room of an ME system
for MRT).
The emission characteristics of this product make it suitable for use in industrial areas as
well as in hospitals (CISPR 11 Class A) and other professional healthcare environments.
If it is used in a residential environment (for which CISPR 11 Class B is normally
required), the product may not offer sufficient protection for radio transmission operation.
The user might need to take mitigation measures, such as relocating or re-orienting the
product.
8.2 Accessories and cables
Accessories and cables for EMC compliance
Type Shield Length [m] Ferrite Use
External adaptor Yes 1.9 ??? Power supply
USB cable Yes 0.2 ??? USB connection
DisplayPort ca-
ble
Yes 0.2 ??? DisplayPort con-
nection
Power supply
unit
Yes 0.2 ??? Power supply
8.3 Table 1 – Compliance level for immunity tests
Guidelines and manufacturer’s declaration– electromagnetic immunity
The product is intended for use in the electromagnetic environment specified below. The user
of the product should make sure that it is used in such an environment.