Electromagnetic compatibility
Instructions for use • OR1 SmartScreen • PQJ219_EN_V1.0_04-2021_IFU_CE-MDR 20
8.4 Table 2 – Test levels for proximity fields from HF
wireless communications equipment
Test fre-
quency
MHz
Frequency
band
MHz
Radio service Modulation Immunity
test level
V/m
Compliance
level
V/m
385 380 – 390 TETRA 400 Pulse modula-
tion 18Hz
27 27
450 430 – 470 GMRS 460,
FRS 460
FM ±5kHz
deviation
1kHz sine
wave
28 28
710 704 – 787 LTE band 13 &
17
Pulse modula-
tion
217Hz
9 9
745
780
810 800 – 960 GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE band 5
Pulse modula-
tion
18Hz
28 28
870
930
1,720 1,700 – 1,990 GSM 1800,
CDMA 1900,
GSM 1900,
DECT,
LTE band 1, 3,
4, 25,
UMTS
Pulse modula-
tion
217Hz
28 28
1,845
1,970
2,450 2,400 – 2,570 Bluetooth,
WLAN 802.11
b/g/n,
RFID 2450,
LTE band 7
Pulse modula-
tion
217Hz
28 28
5,240 5,100 – 5,800 WLAN 802.11
a/n
Pulse modula-
tion
217Hz
9 9
5,500
5,785
8.5 Table 3 – Test levels for radiated and conducted
immunity tests
Guidelines and manufacturer’s declaration– electromagnetic immunity
The product is intended for use in the electromagnetic environment specified below. The user
of the product should make sure that it is used in such an environment.
Immunity tests EN/IEC60601 test
level
Compliance
level
Electromagnetic envi-
ronment – guidelines
Conducted RF distur-
bances acc. to IEC
61000-4-6
3V
rms
150kHz to 80MHz
3V
rms
Portable and mobile HF
communications equip-
ment should be used no
closer to any part of the