ESR-Auto Plus Operator’s Manual 55
Sound pressure level
Less than 55 dB at the highest sound pressure level.
Working environment
Temperature range 5° C to 40° C
Relative humidity Max. 80% for temperatures up to 31° C, decreasing linearly to 50% relative humidity
at 40° C.
Storing environment
Temperature range 0° C to 40° C
Relative humidity Max. 80% for temperatures up to 31° C, decreasing linearly to 50% relative humidity
at 40° C.
Ventilation requirements
Not specied, but one must be aware that the ESR reaction increases with higher temperature.
Dimensions
272mm (width) x 273mm (depth) x 171mm (height)
Weight
10.75 pounds (4.88 kg)
8.11 Environmental Conditions
ESR-Auto Plus is designed to be safe under the following conditions:
a. Indoor use
b. Altitude up to 2,000 meters
c. Temperature 5° C to 40° C
d. Maximum relative humidity 80% for temperatures up to 31° C, decreasing linearly to 50% relative humidity at
40° C
e. Main supply voltage uctuations not to exceed +10% of the nominal voltage
f. Transient over voltages according to installation category II
8.12 Transportation Environment
Temperature Range: -40º C to 55º C.