ESR-Auto Plus Operator’s Manual 71
11.9 Re-Register Test Rack
1. From standby, press . Enter “6.”
2. Instrument reads: “User Cong.” Press .
3. Instrument reads: “Service Menu.” Press .
4. Instrument reads: “Give Code.” Enter “48.” Press . Press
until the instrument reads “Reg Test Rack?”
5. Instrument reads: “Reg Test Rack?” Press .
6. Instrument reads: “Sure?” Press . Instrument reads: "Insert
Rack [OK]." Insert Rack. Press . The instrument will scan
twice and print results. The printout should read “TEST RACK
REGISTRATION OK.”
7. After registration, a test rack verication will occur. Press to
verify the results. The printout should read "SCANNER TEST
OK."
8. Instrument reads: “Set Tube Type?” Press .
9. Instrument reads: “Set Calibration?” Press .
10. Instrument reads: “Set Test Mode?” Press .
11. Instrument reads: “Set Access Code?” Press .
12. Instrument reads: “See Temperature?” Press .
13. Instrument reads: “Next Sample…”
The instrument is ready to run samples.