GENERAL DESCRIPTION
The meter is a handheld 31/2 digital multimeter for measuring DC and AC voltage, DC current,
Resistance, Diode and Transisto. This multimeter is powered by a 9V battery (included)
DURING USE
Never exceed the protection limit values indicated in specications for each range of
measurement
When the meter is linked to measurement circuit, do not touch unused terminals.
Never use the meter to measure voltages that might exceed 600V above earth ground in
category II installations.
When the value sclae to be measured is unknown beforehand, set the range selector at the
highest position .
Before rotating the range selector to change functions , disconnect test leads from the circuit
under test.
When carrying out measurements on TV or switching power circuits always remember that there
may be high amplitude voltages pulses at test points. which can damage the meter.
Always be careful when working with voltages above 60V de or 30V ac rms . Keep ngers behind
the probe barriers while measuring.
Before attempting to insert transistors for testing. always be sure that test leads have been
disconnected from any measurement circuits.
Components should not be connected to the hFE socket when making voltage measurements
with test leads.
Never perform resistance measurements on live circuits.
5