EasyManua.ls Logo

STRILING Shuttle ULT25NE - 4.3 Using Preset Temperatures; 4.4 Over-Temperature Alarm

STRILING Shuttle ULT25NE
28 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
16 of 28
Operating Manual: ULT25NE 10327.1 | 12.22.2016
4.3 – Using Preset Temperatures
• The Temperature Presets — 1, 2 and 3 can be used to quickly change the Setpoint to
frequently used values. When shipped the Presets are set to -86
º
C, -40
º
C, and -20
º
C.
• To choose one of the preset temperatures:
— Press and hold the Set Temp button until the Set Temp LED glows.
— Push one of the Temperature Preset buttons briefly.
— The new Set Temperature will be shown in the LCD display.
— If this is the desired Set Temperature, the user can
either:
– Briefly press the Set Temp button to turn off the Set Temp LED.
— OR
– Wait for approximately eight seconds when the freezer control panel
will return to normal operation.
4.4 – Over-Temperature Alarm
• If the freezer compartment temperature is 10
º
C warmer or colder than the Setpoint
for five seconds then it will go into temperature alarm mode.
— The Over Temp LED will turn on (this is used for under temperature alarms as well).
— The freezer will make a double beep every 10 seconds.
— Situations which can affect the freezer’s ability to maintain a steady state
or cooling rate include:
– The lid is ajar
– The air vents are blocked
– Addition of a large relatively warm mass
– Power supply problems
– Mechanical and/or electrical problems
4.4.1 – How to Silence the Audible Temperature Alarm
• The Over Temp LED will always be lit when the freezer’s temperature is over or under
the set temperature by more than 10
º
C.
4.4.2 – To Postpone an Alarm
• When the audible alarm is sounding it can be postponed.
• Press the Temperature Preset - 3 button (located directly beneath the Over Temp LED)
and the audible alarm will be turned off for 1 hour.