EasyManuals Logo

Stryker CrossFlow Service Manual

Stryker CrossFlow
48 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #30 background imageLoading...
Page #30 background image
EN-28
DRAFT
Error Message Cause Possible Solution
FAULT Instrument recognition error: The stopcock
on the cannula is closed, the lower clamp
on inow tubeset is closed, or no hardware
is attached to the tubing
Check the inow tubing
from the bags to the joint
and ensure the hardware is
properly attached�
Open any closed clamps or
stopcocks�
Repeat the priming step
FAULT Inow and/or Outow Cassette not fully
inserted�
Remove and reinsert the
cassette until it clicks into
place�
FAULT Cassette is expired (after 24 hours of use)� Replace the cassette�
FAULT The connection to the Crossre console or
approved resection console is lost�
Ensure the consoles are
properly connected to the
pump� Follow the instructions
in the “Setup and Device
Conguration section in this
manual�
Reboot the system�
If the problem persists, contact
your Stryker representative or
return the console for repair
FAULT The Crossre console is specied in the
user preference le, but it is improperly or
not connected to the pump
Ensure the console is properly
connected to the pump� Follow
the instructions in the “Setup
and Device Conguration
section in this manual�
Reboot the system�
If the problem persists, contact
your Stryker representative or
return the console for repair
LOCKDOWN Hardware fault detected Reboot the system�
If the problem persists, contact
your Stryker representative or
return the console for repair
LOCKDOWN Motor defective Reboot the system�
If the problem persists, contact
your Stryker representative or
return the console for repair
LOCKDOWN Pressure Transducer out of calibration:
If the pressure is at a value greater than
20mmHg for ve seconds when no
cassette is inserted�
Reboot the system�
If problem persists, contact
your Stryker representative�
LOCKDOWN Possible Overpressure: The pressure is
sustained at 750mmHg for more than one
second, or a hardware failure or pressure
transducer malfunction has occurred�
Reboot the system�
If the problem persists, contact
your Stryker representative or
return the console for repair

Other manuals for Stryker CrossFlow

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker CrossFlow and is the answer not in the manual?

Stryker CrossFlow Specifications

General IconGeneral
BrandStryker
ModelCrossFlow
CategoryWater Pump
LanguageEnglish

Related product manuals