EN-28
DRAFT
Error Message Cause Possible Solution
FAULT Instrument recognition error: The stopcock
on the cannula is closed, the lower clamp
on inow tubeset is closed, or no hardware
is attached to the tubing
Check the inow tubing
from the bags to the joint
and ensure the hardware is
properly attached�
Open any closed clamps or
stopcocks�
Repeat the priming step
FAULT Inow and/or Outow Cassette not fully
inserted�
Remove and reinsert the
cassette until it clicks into
place�
FAULT Cassette is expired (after 24 hours of use)� Replace the cassette�
FAULT The connection to the Crossre console or
approved resection console is lost�
Ensure the consoles are
properly connected to the
pump� Follow the instructions
in the “Setup and Device
Conguration section in this
manual�
Reboot the system�
If the problem persists, contact
your Stryker representative or
return the console for repair
FAULT The Crossre console is specied in the
user preference le, but it is improperly or
not connected to the pump
Ensure the console is properly
connected to the pump� Follow
the instructions in the “Setup
and Device Conguration
section in this manual�
Reboot the system�
If the problem persists, contact
your Stryker representative or
return the console for repair
LOCKDOWN Hardware fault detected Reboot the system�
If the problem persists, contact
your Stryker representative or
return the console for repair
LOCKDOWN Motor defective Reboot the system�
If the problem persists, contact
your Stryker representative or
return the console for repair
LOCKDOWN Pressure Transducer out of calibration:
If the pressure is at a value greater than
20mmHg for ve seconds when no
cassette is inserted�
Reboot the system�
If problem persists, contact
your Stryker representative�
LOCKDOWN Possible Overpressure: The pressure is
sustained at 750mmHg for more than one
second, or a hardware failure or pressure
transducer malfunction has occurred�
Reboot the system�
If the problem persists, contact
your Stryker representative or
return the console for repair