EN-36
DRAFT
Technical Specications
Equipment Information
Size Pump Dimensions:
• 12�528in� (318�2mm) width
• 7�025in� (178�4mm) height
• 16�990in� (431�6mm) depth
Pump Weight: 24lbs� (10�9kg)
Hospital power cord: (p/n 0105-033-001) 2m
Approved resection console power cord: (p/n P17275) 2m
Power
Connection
Input Voltage: 80–275V
Inlet Fuse: 10A, 250V
Frequency 47–63Hz
Power/Current
Consumption
Maximum power consumption: 269W
Maximum current consumption: 80V: 3�4A; 275V: 0�98A
Motor output maximum speed: 650RPM
RF output waveform: 13�56MHz (ISO15693)
Electrical Safety
Classication
Class I equipment
Type BF Part
Water ingress protection, IPX1
Continuous Operation
Environmental
Specications
Operating Temperature: 5 – 40°C
Operating Humidity: 30 – 95% RH (non-condensing)
Shipping Temperature: -18° – 60°C
Shipping Humidity: 15 – 90% RH
Safety and EMC
Compliance
• UL 60601-1: 2006
• IEC 60601-1-2: 2007
• IEC 60601-1-4: 2000
• IEC 60601-1-6: 2010
• IEC 60601-1-8: 2006
• IEC 62304: 2006
• IEC 62366: 2007
• CAN/CSA-C 22�2 No� 601�1-M90: 2003
• CAN/CSA 22�2 No� 60601-1: 2002
• AS/NZS 3200�1�0: 1998
• IEC 60601-1: 2005 + Corr 2006 + Corr 2007
• IEC 60601-1-1: 2005 + Corr 2006 + Corr 2007
• CAN/CSA-C 22�2 No� 60601-1-2: 2003 + A1: 2006
Pump Capacity 3�0L/min
Pressure range 0-150mmHg
Max suction by
pressure relief
500mmHg
Measuring
accuracy
Pressure: ≤2%
Flow: ±10%
Display/Color
Touch Screen
160° viewing angle
6�5inch diagonal active LCD
Resolution 800 (horizontal) x 480 (vertical)
12:8 aspect ratio
16-bit color
Dimensions: 153mm (width) x 118mm (height)
Volume
Adjustment
0-85dBA