EasyManua.ls Logo

Stryker CrossFlow - Technical Specifications; Equipment Information Details

Stryker CrossFlow
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
EN-36
DRAFT
Technical Specications
Equipment Information
Size Pump Dimensions:
12�528in� (318�2mm) width
7�025in� (178�4mm) height
16�990in� (431�6mm) depth
Pump Weight: 24lbs� (10�9kg)
Hospital power cord: (p/n 0105-033-001) 2m
Approved resection console power cord: (p/n P17275) 2m
Power
Connection
Input Voltage: 80–275V
Inlet Fuse: 10A, 250V
Frequency 47–63Hz
Power/Current
Consumption
Maximum power consumption: 269W
Maximum current consumption: 80V: 3�4A; 275V: 0�98A
Motor output maximum speed: 650RPM
RF output waveform: 13�56MHz (ISO15693)
Electrical Safety
Classication
Class I equipment
Type BF Part
Water ingress protection, IPX1
Continuous Operation
Environmental
Specications
Operating Temperature: 5 – 40°C
Operating Humidity: 30 – 95% RH (non-condensing)
Shipping Temperature: -18° – 60°C
Shipping Humidity: 15 – 90% RH
Safety and EMC
Compliance
UL 60601-1: 2006
IEC 60601-1-2: 2007
IEC 60601-1-4: 2000
IEC 60601-1-6: 2010
IEC 60601-1-8: 2006
IEC 62304: 2006
IEC 62366: 2007
CAN/CSA-C 22�2 No� 601�1-M90: 2003
CAN/CSA 22�2 No� 60601-1: 2002
AS/NZS 3200�1�0: 1998
IEC 60601-1: 2005 + Corr 2006 + Corr 2007
IEC 60601-1-1: 2005 + Corr 2006 + Corr 2007
CAN/CSA-C 22�2 No� 60601-1-2: 2003 + A1: 2006
Pump Capacity 3�0L/min
Pressure range 0-150mmHg
Max suction by
pressure relief
500mmHg
Measuring
accuracy
Pressure: ≤2%
Flow: ±10%
Display/Color
Touch Screen
160° viewing angle
6�5inch diagonal active LCD
Resolution 800 (horizontal) x 480 (vertical)
12:8 aspect ratio
16-bit color
Dimensions: 153mm (width) x 118mm (height)
Volume
Adjustment
0-85dBA

Other manuals for Stryker CrossFlow

Related product manuals