EN-42
DRAFT
Icon Name/Term Description
45
mmHg
PRESSURE Displays the current pressure setting
FLOW/SUCTION Displays the current ow/suction setting
START/STOP Starts/stops the pump
IN PROGRESS Indicates the pump is operating
HOT SWAP Toggles between hardware combination selections
WASH FUNCTION Increases the set pressure and ow limit by a user-specied
percentage over a user-specied duration (for Inow-only
mode)
Increases the set pressure and suction by a user-specied
percentage over a user-specied duration (for Inow/Outow
mode)
DRAIN FUNCTION Removes excess uid from the joint at the end of a case
(Outow mode only)
HOME Retuns to the user preference le selection screen
DELETE Deletes a le
CANCEL Rejects an action
CONFIRM Accepts an action
HELP Accesses task-oriented help related to screen and button
functionality
WASH MENU Navigates to the Wash menu to adjust the Wash and Clear
function settings
CLEAR FUNCTION Increases the ow limit by a user-specied percentage over a
user-specied duration (for Inow-only mode)
Increases the suction by a user-specied percentage over a
user-specied duration (for Inow/Outow mode)
RESECTION INTEGRATION Navigates to resection integration settings menu
RF PROBE Selects an approved resection RF console
SHAVER Selects an approved resection shaver console
CROSSFIRE Selects the Stryker Crossre console
APPROVED RESECTION
CONSOLE
Selects an approved resection console
NONE Indicates no RF/shaver console is selected
FOOTSWITCH/FORMULA
SHAVER OPTIONS
Programs footswitch and hand control options for Crossow,
iSwitch, Crossre, and Formula shaver