Summary of Safety Precautions
Pump
− Risk of Electric Shock. Do not open or attempt to repair or service the electronic Pump. Repairs and service
should only be done by authorized personnel. If the Pump is not functioning properly, or has been damaged,
unplug the Pump and take it out of service immediately and contact Customer Service. (See Page 9 for
contact information).
− Electrical-safety testing of your Pump should be performed at least annually. Failure to do so may result in
death or injury. Contact Customer Service, Page 9, for service information.
− This equipment radiates radio frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to other devices in the vicinity (See Pages 45-46). However,
there is no guarantee that interference will not occur in a particular installation. If this equipment does cause
harmful interference to other devices, which can be determined by turning the equipment ON and OFF, the
user is encouraged to try to correct the interference by one or more of the following measures:
1. Reorient or relocate the receiving device.
2. Increase the separation between the Pump and other equipment.
3. Connect the equipment into an outlet on a circuit different from that to which other device(s) are
connected.
Consult with Stryker Customer Service for assistance.
− Wireless communication equipment, such as wireless home network devices, mobile phones, cordless
telephones and their base station, walkie-talkies can affect this equipment. Place equipment a minimum
distance from wireless communication devices as defined on Page 46.
− The use of accessories and cables other than those specified, with the exception of those sold by the
manufacturer as replacement parts, may result in increased emissions and/or decreased immunity of the
device. (See Pages 45-46).
− The Pump hangers are not intended to be in patient contact. Extended patient contact with the Pump hangers
may cause injury.
− Do not use in the presence of flammable anesthetics, nitrous oxide, or oxygen-rich environments. Risk of
explosion, burns and asphyxiation can result.
− Exposure of the electronic Pump to any liquid while it is plugged in could result in a severe electrical hazard.
− To avoid risk of injury do not place objects on the surface of the Pump.
− The AST cable ONLY connects to the AST socket. Connecting it anywhere else may result in severe
electrical shock.
− If “Key-Click” sound is not heard, DO NOT use the Pump.
− Pressure in support surface is under automated control and may adjust without notice. Use care when
performing medical procedures on patient.
− The device is not compatible for use in MRI.
− AC mains power must be connected to provide therapy. If power is lost, therapy provided will be discontinued.
− Good filter maintenance is critical in keeping your IsoAir™ Pump in optimal operating condition. Failure to
clean the filter may cause damage to the Pump. A damaged Pump may not provide proper support pressures
resulting in patient injury.