Stryker GoBed II - Service Manual

Stryker GoBed II
338 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
GoBed
®
II MedSurg Bed
GoBed
®
II Lit MedSurg
Model FL28C / Modèle FL28C
Maintenance Manual
2011/11 B.6 280C-009-002 REV B www.stryker.com
For Parts or Technical Assistance:
USA: 1-800-327-0770
Guide d’entretien

Table of Contents

Questions and Answers

  • V
    Vicki BurtonAug 19, 2025
    How to troubleshoot Stryker GoBed II Medical Equipment bed with no power?
    • P
      Phillip BennettAug 19, 2025
      If your Stryker Medical Equipment bed has no power, verify the power at the wall receptacle and the power cord connections at both the wall and the bed. Inspect the power cord for any damage and replace it if needed. Also, check the two fuses inside the power connector.
  • S
    scottthompsonAug 21, 2025
    What to do if Stryker Medical Equipment scale displays 'OVERLOAD' or incorrect weight?
    • W
      wigginsangelicaAug 21, 2025
      If the scale display on your Stryker Medical Equipment shows 'OVERLOAD' or an incorrect weight, first ensure the patient's weight is not greater than 500 lb (227 kg). If the weight is within the limit, zero the scale without the patient on the bed. Also, make sure any equipment added to the bed was done following the 'Add/Remove Equipment Procedure'. Finally, check that all load cell cables are properly connected to the scale control board.
  • K
    Kayla SilvaAug 25, 2025
    Why is there no bed up or down motion on my Stryker GoBed II Medical Equipment?
    • L
      leeclarkAug 25, 2025
      If your Stryker Medical Equipment bed isn't moving up or down, ensure that the Total Lockout feature (padlock LED) is not enabled. Check the siderail control panel cable connection to the bed receptacle located under the mattress support. Also, verify that the bed has power.
  • M
    Michele WilsonAug 26, 2025
    What to do if Fowler is not moving up or down on Stryker Medical Equipment?
    • S
      Sandra FrancoAug 26, 2025
      If the Fowler function on your Stryker Medical Equipment bed isn't moving up or down, make sure that neither the Total Lockout nor the Fowler Lockout (padlock LEDs) are enabled. Check the siderail control panel cable connection to the bed receptacle, which is under the mattress support. Also, ensure that the bed has power.
  • A
    Andre ThompsonAug 28, 2025
    Why is there no Gatch up or down motion on my Stryker GoBed II bed?
    • J
      jonathan45Aug 28, 2025
      If the Gatch function on your Stryker Medical Equipment bed isn't moving up or down, ensure that neither the Total Lockout nor the Gatch Lockout (padlock LEDs) are enabled. Check the siderail control panel cable connection to the bed receptacle located under the mattress support. Also, verify the bed has power.
  • D
    Deanna FischerAug 30, 2025
    How to fix Stryker Medical Equipment scale display that doesn't turn on?
    • L
      lindasummersAug 30, 2025
      If the scale display on your Stryker Medical Equipment does not turn on when activated, first verify that the bed has power. Then, press the ENTER key twice. Unhook the foot board from the bed by lifting it off the connectors, ensure the connectors match correctly, and press the foot board down firmly to reconnect it to the bed. Finally, check the Scale user interface cable connection to the scale control board (J6 connector).
  • S
    Sean LiuSep 1, 2025
    How to troubleshoot no auto contour motion on Stryker GoBed II?
    • W
      William WilliamsSep 1, 2025
      If the Auto Contour motion is not working on your Stryker Medical Equipment bed, ensure that the Gatch Lockout (padlock LED) is not enabled. Check the two Auto Contour limit switches and replace them if needed.
  • R
    Rachel JonesSep 3, 2025
    What to do if Stryker GoBed II bed exit system doesn't work or alarm goes off?
    • R
      Ralph MillerSep 3, 2025
      If the Bed Exit system on your Stryker Medical Equipment is not working or the alarm goes off when activated, check the load cell cable connections to the scale control board.
  • K
    Kirk ColeSep 5, 2025
    How to fix Stryker GoBed II Medical Equipment CPR positioning malfunction?
    • S
      santiagojoseSep 5, 2025
      If the CPR positioning is not operating properly on your Stryker Medical Equipment, with the Gatch not lowering and/or the Fowler actuator not resetting, check the two CPR limit switches and replace them if necessary.
  • J
    John ThompsonSep 8, 2025
    Why doesn't the Fowler on my Stryker Medical Equipment reach full height?
    • S
      Samantha HaleySep 8, 2025
      If the Fowler on your Stryker Medical Equipment does not reach full height, it might be because the CPR handle was used to partially lower it. This can cause the Fowler motor to lose sync with the actual position. To fix this, completely lower the Fowler using either the CPR handle or the Fowler down control to allow the motor to reset itself.

Summary

Symbols and Definitions

Warning; Caution; Note Definition

Explains the meaning of warning, caution, and note symbols used in the manual.

Introduction

Intended Use of Product

Defines the intended purpose and environment for the GoBed II MedSurg Bed, Model FL28C.

Specifications

Details the bed's technical characteristics, including load capacity and dimensions.

Environmental Conditions

Specifies the operating and storage conditions for the bed.

Safety Tips and Guidelines

General Safety Precautions

Provides essential safety rules for operating and maintaining the bed.

Specific Hazard Warnings

Highlights potential risks like electrical hazards, patient injury, and proper usage.

Static Discharge Precautions

Static Protection Procedure

Outlines the steps for grounding oneself before servicing electronics.

Cleaning

Approved Cleaning Agents

Lists recommended cleaners and cautions for cleaning the bed surfaces.

Preventative Maintenance

Maintenance Checklist

A comprehensive list of items to inspect annually for bed maintenance.

Lubrication Requirements

Details the lubrication schedule and procedure for actuator screws.

Troubleshooting Guide

Common Problems and Recommended Actions

Lists typical bed malfunctions and their solutions.

Quick Reference Replacement Parts List

Electric;Electronic Components

Lists part names and numbers for electrical and electronic components.

Base Components

Lists part names and numbers for base structural components.

Electrical Service Information

Motor Control Board

Provides a wiring diagram for the motor control board.

Power Supply

Details the electrical system diagrams for different power supply configurations.

Scale System (Optional)

Wiring diagram for the optional scale system.

Maintenance Menu: Error Codes

Explains error codes related to the scale system.

Service Information

Head End Siderail Assembly Replacement

Step-by-step guide to replace the head end siderail assembly.

Foot Section Replacement

Procedure for replacing the foot section of the bed.

Fowler Motor Replacement

Instructions for replacing the fowler motor.

Control Board Replacement

Steps for replacing the main control board.

Brake System

Brake System Components

Lists the parts and quantities for the brake system.

Lift Mechanism

Lift Mechanism Components

Lists the parts and quantities for the lift mechanism.

Optional Accessories

Pendant Assemblies

Details the various pendant control options available for the bed.

Monitor Tray

Lists parts for the optional monitor tray accessory.

Patient Helper

Lists parts for the optional patient helper accessory.

Oxygen Bottle Holder

Lists parts for the optional oxygen bottle holder accessory.

Bed Extender

Lists parts for the optional bed extender accessory.

I.V. Poles

Lists parts for the removable and fixed I.V. poles.

EMC Information

Electromagnetic Immunity Guidance

Provides guidance on electromagnetic immunity based on IEC 60601 standards.

RF Equipment Separation Distances

Recommends distances for RF equipment to prevent interference.

Warranty

Limited Warranty

Outlines the terms and conditions of the product's limited warranty.

Service Contract Programs

Details available service contract options for equipment maintenance.

Stryker GoBed II Specifications

General IconGeneral
Weight Capacity500 lbs (227 kg)
Mattress Length80 in (203 cm)
Frame MaterialSteel
Backrest Angle Range0-70 degrees
Knee Gatch Range0-30 degrees
Wheel Diameter5 in
Reverse Trendelenburg Range30 degrees
Power SourceAC/DC
FeaturesAdjustable height, Trendelenburg positioning