EasyManuals Logo

Stryker GoBed II Service Manual

Stryker GoBed II
338 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #164 background imageLoading...
Page #164 background image
1-162 280C-009-002 REV B www.stryker.com
Return To Table of Contents
English
Guidance and Manufacturer’s declaration - Electromagnetic Immunity
The GoBed
®
II MedSurg Bed, Model FL28C is suitable for use in the electromagnetic environment specified below.
The customer or the user of the GoBed
®
II MedSurg Bed, Model FL28C should assure that it is used in such an
environment.
Immunity Test IEC 60601 Test Level Compliance Level
Electromagnetic
Environment Guidance
Electrostatic Discharge (ESD)
IEC 61000-4-2
+6 kV contact
+8 kV air
+6 kV contact
+8 kV air
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with synthetic
material, the relative humidity
should be at least 30%.
Electrostatic fast
Transient/burst
IEC 61000-4-4
+2 kV for power
supply lines
+1 kV for input/
output lines
+2 kV for power
supply lines
+1 kV for input/
output lines
Main power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
+8 kV differential mode
+2 kV common mode
+8 kV differential mode
+2 kV common mode
Main power quality is that of
a typical commercial and/or
hospital environment.
Voltage dips, voltage variations
and short interruptions on
power supply input lines
IEC 61000-4-11
<5%Ut (95% dip in Ut) for
0.5 cycle
40%Ut (60% dip in Ut) for
5 cycles
70%Ut (30% dip in Ut) for
25 cycles.
<5% Ut (>95% dip in Ut)
for 5 sec.
<5%Ut (95% dip in Ut)
for 0.5 cycle
40%Ut (60% dip in Ut)
for 5 cycles
70%Ut (30% dip in Ut)
for 25 cycles.
<5% Ut (>95% dip in Ut)
for 5 sec.
Main power quality should be
that of a typical commercial
and/or hospital environment.
If the user of the GoBed
®
II
MedSurg Bed, Model FL28C
requires continued operation
during power main interruptions,
it is recommended that the
device be powered from an
uninterrupted power supply or
a battery.
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
and/or hospital environment.
Note: U
T
is the a.c. mains voltage prior to applications of the test level.
GOBED
®
II MEDSURG BED, MODEL FL28C
EMC Information

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker GoBed II and is the answer not in the manual?

Stryker GoBed II Specifications

General IconGeneral
BrandStryker
ModelGoBed II
CategoryMedical Equipment
LanguageEnglish

Related product manuals