EasyManuals Logo

Stryker GoBed II Service Manual

Stryker GoBed II
338 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #62 background imageLoading...
Page #62 background image
1-60 280C-009-002 REV B www.stryker.com
Return To Table of Contents
English
CPR ACTIVATION CABLE REPLACEMENT (OPTIONAL)
Tools Required:
• 7/16” Combination Wrench
Note
Unless otherwise indicated, refer to Figure 17, page 1-59, for this procedure.
Procedure:
1. Raise the bed to the highest position and apply the brakes.
2. Remove the head section plastic cover (if equipped) or raise the head section completely.
3. Remove the head board.
4. Unplug the power cord from the wall receptacle.
5. Using a 7/16” combination wrench, remove the nut, sleeve and bolt (D) holding both ends of the activation cable to
the fixed lever.
6. Using a 7/16” combination wrench, remove the nut, washer (E) holding the cable jacket to the mobile lever. Remove
the cable from the lever and keep the spring (F). Use caution to avoid moving the other cable from its position.
7. Unscrew the two lock nuts (G) holding the jacket of the other end of the cable to the CPR handle. Remove the
wire from the handle. Once the new cable is installed, the free play of the CPR handle must not be more than 1/8”.
8. Reverse the above steps to install the new cable.
9. Test the CPR positioning before returning the bed to service.
Service Information

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker GoBed II and is the answer not in the manual?

Stryker GoBed II Specifications

General IconGeneral
BrandStryker
ModelGoBed II
CategoryMedical Equipment
LanguageEnglish

Related product manuals