1-50 280C-009-002 REV B www.stryker.com
Return To Table of Contents
English
POWER CORD CONNECTOR FUSE REPLACEMENT
Tools Required:
• Small Regular Screwdriver
Procedure:
Note
Unless otherwise indicated, refer to Figure 12, page 1-48, for reference for this procedure.
1. Remove the power cord from the power cord connector.
2. Using a small regular screwdriver, remove the fuse holder.
3. Remove the defective fuse and replace it with a new one.
Note
120 VAC rated FL28C beds use fast acting, 250 Vac, 10A type fuses. For beds rated at voltages other than 120 Vac, see
“Specifications”, page 1-6, for fuse ratings.
4. Replace the fuse holder.
Note
Make sure to use the right fuse holder for the fuses used with the bed and make sure the fuses are correctly installed
to be stable in the fuse holder.
5. Test the bed for proper operation before returning the bed to service.