How to troubleshoot no power to Stryker Medical Equipment bed?
- PPeter AndrewsAug 3, 2025
If your Stryker Medical Equipment bed has no power, start by verifying the power cord connections at both the wall and the bed. Next, check the circuit breakers located under the litter/gatch section on the patient's left side; if a breaker is tripped, reset it by pushing it in. If these steps don't restore power, check for 120 VAC at J1 on the power supply, Pin 1 and 2, and check for DC voltages on J2 (Pins 1, 2, 3 & 6) on the power supply. If voltage is not present, replace the power supply. If DC voltages do not come back, replace the CPU board. Also, check for 120 VAC at connector O on the CPU board. If voltage is present, replace the CPU board. Finally, verify bed function and return to service.