Zoomr Option Inverter Protection Features
57
The inverter has several features to prevent internal damage:
1. Low Battery Voltage If the battery voltage at the inverter drops below the low voltage cutoff, the inverter
will shut off.
2. OverTemperature If the inverter gets too hot, it will shut off. The overheating may be caused by high
ambient temperature, blocked air flow or an overload condition. When the inverter reaches an acceptable
temperature, it will restart.
3. OverPower The inverter will source up to its maximum power rating. If the load requires more, the
output voltage will shut down. Turning the power switch off and on will reset the inverter. Plugging the
bed power cord into the wall socket to charge the battery will reset the inverter.
WARNING
The inverter generates 115VAC, the same as a wall receptacle. To prevent injury, do not put anything into
the electrical outlets other than an appliance power cord. Keep the outlets covered when not in use. Do not
submerge the unit or subject it to moisture.