Set−Up Procedures
10
SET−UP PROCEDURES
It is important that the Secure II Bed is working properly before it is put into service. The following list will
help ensure that each part of the bed is tested.
Plug the bed into a properly grounded, hospital grade wall receptacle and ensure the ”Power” LED light
at the foot end of the bed comes on.
WARNING
The Secure II is equipped with a hospital grade plug for protection against shock hazard. It must be plugged
directly into a properly grounded three−prong receptacle. Grounding reliability can be achieved only when
a hospital grade receptacle is used.
Plug the optional interface cable into the 37 pin connector under the litter frame at the head end of the
bed, and into the ”Patient Station”, ”Head Wall”, ”Docker Station”, or equivalent (whichever applies). Test
the interface cable to verify it is functioning properly.
WARNING
Use only a Stryker supplied interface cable. Use of any other cable may cause the bed to function improperly
which may result in patient or user injury.
Ensure the siderails raise, lower and store smoothly and lock in the up position and in the intermediate
position when lowered (page 18).
Ensure that all four casters lock when the brake pedal is engaged (page 14).
Raise the Back up to approximately 60. Squeeze the CPR release handle and ensure the Back will drop
with minimal effort.
NOTE
Ensure that the ”Brake Not Set” LED’s located on the outside of the head end siderails and on the foot board
control panel come on when the brakes are disengaged.
Run through each function on the foot board control panel to ensure that each function is working properly
(page 21−25).
Run through each function on both head end siderails to ensure that each is working properly (page
18−20).
Activate the motion stop system to ensure it is functioning properly: press and hold down the BED DOWN
key. As the bed lowers, lift up on the motion interrupt pan (reference drawing on page 13) and ensure
the downward motion stops. Release the pan and allow the downward motion to continue.
NOTE
The bed’s upward motion or other functions are not disrupted by the motion stop system.