Maintenance Procedures Siderail
90
HEAD AND FOOT SIDERAIL COVER REMOVAL
Required Tools:
#2 Phillips Screwdriver
Removal Procedure:
1. Unplug the power cord from the wall receptacle.
2. Using a #2 Phillips screwdriver, remove the 8 phillips screws (A) holding the covers (B) to the siderail.
CAUTION
There are two cables connecting the outside cover to the head end siderail. Be careful not to pull on them
when removing the cover.
3. Remove the cables from the siderail. Make note of the proper location for the cables.
4. Reverse the above steps to reattach the cover.
CAUTION
Do not snag the cables when installing the siderail cover.
NOTE
Follow the same procedure for siderail cover removal for the foot end rails.