Service Information
61
LIFT MOTOR AND CAPACITOR REMOVAL AND REPLACEMENT
Required Tools:
3/8” Socket Wrench w/Extension 5/16” Socket Wrench Floor Jack
Side Cutters 7/16” Open End Wrench 2x4 (or Equivalent)
A
B
D
FOOT END
C
Procedure:
NOTE
If you need more space to work under the base frame, place a 2x4 across the base frame rails and use a
floor jack to raise the base frame off the floor.
1. Unplug the bed power cord from the wall socket. Using a 5/16” socket wrench, remove the five bolts
holding the lower lift cover to the base and remove the cover.
2. Disconnect the two connectors (A) at the motor capacitor.
3. Disconnect the white connector (B) from the power cord.
4. Using side cutters, cut the cable ties holding the capacitor (C) to the base and remove the capacitor.
5. Using a 3/8” socket wrench, remove the four screws (D) holding the motor assembly in the lift housing
and remove the motor assembly.
6. Reverse the above steps to install the new motor.
NOTE
The drive shaft on the new motor probably will have to be turned to be aligned with the coupler. Use a 7/16”
open end wrench to turn the drive shaft of the motor.
The procedure for lift motor and capacitor removal and replacement is the same for both ends of the bed.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker Secure II 3002 and is the answer not in the manual?

Stryker Secure II 3002 Specifications

General IconGeneral
BrandStryker
ModelSecure II 3002
CategoryMedical Equipment
LanguageEnglish

Related product manuals