Service Information
76
HEAD MOTOR REMOVAL AND REPLACEMENT
Required Tools:
3/8” Socket Wrench
Procedure:
WARNING
The head motor is not grounded and may be an electrical hazard. Use caution when working around motors.
1. Using the CPR release, lower the Fowler to flat (0).
2. Unplug the bed power cord from the wall socket.
3. Using a 3/8” socket wrench, remove the two bolts at the foot end of the motion interrupt pan and remove
the pan.
4. Using a 3/8” socket wrench, remove the two bolts holding the motor bracket to the bed frame. Remove
the motor.
NOTE
Use caution when removing the motor to avoid having the coupler come off the ball screw shaft.
5. Unplug the head motor cables from the bed and capacitor.
NOTE
The replacement motor will come with the attached CPR clutch. The mounting bracket will be re−used with
the replacement motor.
6. Reverse the above steps to install the new motor assembly and reattach all litter panels.