Maintenance Procedures Siderail
91
HEAD AND FOOT MOLDED SIDERAIL REPLACEMENT
Required Tools:
#2 Phillips Screwdriver 3/8” Nut Driver
Procedure:
1. Unplug the bed power cord from the wall socket.
2. Remove the siderail cover (see page 90).
3. Using a 3/8” nut driver, remove the four screws (A) holding the molded rail (C) to the siderail support
assembly (B).
NOTE
Note the location of the spacers (D) for reassembly purposes.
4. Pull up on the molded rail (C) to remove it from the siderail assembly.
5. Reverse the above steps to install the new molded rail.