Inverter/Charger, Battery & Zoomr Troubleshooting
44
Connect Bed to
Wall Power (~120V AC)
Replace the Inverter
/ Charger assembly
Are the Circuit Breakers OK ?
Reset the
Circuit Breaker
Is the Input supply Voltage
(~120V AC) Good?
Is the Battery Voltage
Increasing?
Is Battery Volage Sufficient?
(> 21VDC)
Replace the Batteries
Battery and Charger Module
No
Yes
No
Yes
No
Yes
Yes
Check Wall Outlet
No
Plug Bed into Wall Socket
(For Charging/Test)
Return Bed to Service