SetUp Procedures
11
SETUP PROCEDURES (ZOOMt OPTION)
If your bed is equipped with the Zoom drive wheel option, run through the setup procedures on page 10
and continue with the procedures listed below.
Plug the power cord into a properly grounded, hospital grade wall receptacle. The 12 volt batteries that
provide power to the drive wheel and backup power to the unit functions will charge whenever the power
cord is plugged into the wall socket. The batteries require approximately 10 hours of charging time before
the bed is put into service.
Unplug the power cord from the wall socket. Push the battery power switch located on the lower left cor-
ner of the head end to the “ON” position. Again, verify each function on the foot board and siderails is
operating properly.
With the battery power switch in the “ON” position and the brakes engaged, ensure the “Release Brakes”
LED on the head end control panel is illuminated.
With the battery power switch in the “ON” position and the drive wheel in the neutral position (not touching
the floor), ensure the “Engage Drive Wheel” LED on the head end control panel is illuminated.
Run through the operation of the drive wheel (see page 33 & 34) to ensure it is operating properly.
If any problems are found during bed setup, contact Stryker Customer Service at 8003270770.
Damaged Merchandise
ICC Regulations require that claims for damaged merchandise must be made with the carrier within fifteen
(15) days of receipt of merchandise. DO NOT ACCEPT DAMAGED SHIPMENTS UNLESS SUCH DAMAGE
IS NOTED ON THE DELIVERY RECEIPT AT THE TIME OF RECEIPT. Stryker Customer Service must be
notified immediately. Stryker will aid the customer in filing a freight claim with the appropriate carrier for dam-
ages incurred. Claim will be limited in amount to the actual replacement cost. In the event that this information
is not received by Stryker within the fifteen (15) day period following the delivery of the merchandise, or the
damage was not noted on the delivery receipt at the time of receipt, the customer will be responsible for pay-
ment of the original invoice in full.
Claims for any short shipment must be made within thirty (30) days of invoice.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker Secure II 3002 and is the answer not in the manual?

Stryker Secure II 3002 Specifications

General IconGeneral
BrandStryker
ModelSecure II 3002
CategoryMedical Equipment
LanguageEnglish

Related product manuals