Gatch Angle
- With Auto Contour (Optional)
- Without Auto Contour
0° to 24°
0° to 32°
Trendelenburg/Reverse Trendelenburg +14° to -1
Electrical Requirements - all electrical requirements meet
CSA C22.2 No. 601.1, UL 60601-1 and IEC 60601-1, 60601-
2-38 specifications.
100Vac, 50-60Hz, 7.5A - Two 250Vac, 10A Fuses
120Vac, 50-60Hz, 4A (9.8A with 120Vac Optional Auxiliary
Outlet) - Two 250V, 10A Fuses
200Vac, 50-60Hz, 3.2A -Two 250Vac, 6.3A Fuses
220Vac, 50-60Hz, 2.9A -Two 250Vac, 6.3A Fuses
240Vac, 50-60Hz, 2.7A -Two 250Vac, 6.3A Fuses
Duty Cycle 10%
Stryker reserves the right to change specifications without notice.
• Specifications listed are approximate and may vary slightly from unit to unit or by power supply fluctuations.
ENVIRONMENTAL CONDITIONS
Environmental Conditions Operation Storage and Transportation
Ambient Temperature
-22
0
F
140
0
F
(-30
0
C)
(60
0
C)
Relative Humidity
(Non-Condensing)
10%
95%
Atmospheric Pressure
Stryker reserves the right to change specifications without notice.
• Specifications listed are approximate and may vary slightly from unit to unit..
• Operating environment recommended to ensure the scale system (optional) precision.
65
0
F
80
0
F
(18
0
C)
(27
0
C)
30%
75%
700 hPa
1060 hPa
500 hPa
1060 hPa