EasyManuals Logo

Stryker Power-PRO XT Maintenance Manual

Stryker Power-PRO XT
186 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #12 background imageLoading...
Page #12 background image
The emissions characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11
class A). If it is used in a residential environment, for which CISPR 11 class B is normally required, this equipment might
not offer adequate protection to radio frequency communication services. The user might need to take mitigation
measures, such as relocating or reorienting the equipment.
Portable RF communications equipment, including peripherals such as antenna cables and external antennas, should
be used no closer than 12 inches (30 cm) to any part of PPoowweerr--PPRROO and SSMMRRTT charger, including cables specified by
the manufacturer.
Avoid stacking or placing other equipment adjacent to PPoowweerr--PPRROO and SSMMRRTT charger to prevent improper operation of
the products. If such use is necessary, carefully observe PPoowweerr--PPRROO and SSMMRRTT charger and the other equipment to
make sure that they are operating properly.
PPiinncchh ppooiinnttss
WWAARRNNIINNGG -- Always keep your hands clear of the red safety bar pivots when you load, unload, or change the height
position of the cot.
FFiigguurree 11 PPiinncchh ppooiinnttss
MMeecchhaanniiccaall ssttaabbiilliittyy
WWAARRNNIINNGG -- Always use both hands when you transport the cot.
EN 6 6506-109-002 Rev G.1

Other manuals for Stryker Power-PRO XT

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker Power-PRO XT and is the answer not in the manual?

Stryker Power-PRO XT Specifications

General IconGeneral
ModelPower-PRO XT
CategoryMedical Equipment
Weight Capacity700 lbs (318 kg)
Power SourceBattery
Battery TypeSealed lead acid
Frame MaterialAluminum
HeightAdjustable
Minimum Height14 in (36 cm)
Wheels4
Incline CapabilityUnspecified
ControlHandheld
Safety FeaturesSide rails

Related product manuals