EasyManuals Logo

Stryker Power-PRO XT Maintenance Manual

Stryker Power-PRO XT
186 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #25 background imageLoading...
Page #25 background image
IItteemm IInnssppeecctt
Litter All welds are intact, not cracked, or broken
Warning labels present, legible
Base All welds are intact, not cracked, or broken
Wheels Check and adjust wheel locks
Retractable head section oxygen bottle holder (optional) Straps and clips for wear
Defibrillator platform (optional) Straps not frayed or torn
Latch hooks are intact and secure
FFoooott eenndd ffaasstteenneerr ppaarrtt rreeppllaacceemmeenntt sscchheedduullee
For PPeerrffoorrmmaannccee--LLOOAADD compatible cots, you must replace the foot end fastener parts every 18,078 calls. This is to make
sure that the PPeerrffoorrmmaannccee--LLOOAADD remains functional. Follow this call volume time table to remain compliant with this
requirement.
CCaallllss ppeerr ddaayy MMoonntthhss
7 Not applicable
8 77
9 67
10 59
MMaaiinntteennaannccee rreeccoorrdd
DDaattee MMaaiinntteennaannccee ooppeerraattiioonn
ppeerrffoorrmmeedd
BByy HHoouurrss
6506-109-002 Rev G.1 19 EN

Other manuals for Stryker Power-PRO XT

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker Power-PRO XT and is the answer not in the manual?

Stryker Power-PRO XT Specifications

General IconGeneral
ModelPower-PRO XT
CategoryMedical Equipment
Weight Capacity700 lbs (318 kg)
Power SourceBattery
Battery TypeSealed lead acid
Frame MaterialAluminum
HeightAdjustable
Minimum Height14 in (36 cm)
Wheels4
Incline CapabilityUnspecified
ControlHandheld
Safety FeaturesSide rails

Related product manuals