OPERATING MANUAL FOR SELECT EDI 60
L300180 Iss:A 17
Set Perm Temp Alarm
BACK
ENTER
30.0
°C
Maintenance
BACK
ENTER
Pre-treat(PT) pack date
Set temperature alarm
If dispensed water must be supplied below a certain temperature an
alarm can be set. Enter the maximum water temperature in the “”Set
Perm Temp alarm” menu. During “Processing” should the water
temperature rise above this, the navigation bar will display
“Temperature High alarm message. If the audible alarm is enabled a
buzzer will sound.
The range of the alarm set point is between 20.0 35.0
o
C. The unit
will be set on delivery to 35.0
o
C
To change the displayed value simply use the
,
,
up/down buttons
to either increase or decrease the value, pressing the NEXT” button
will move the cursor to the next digit. Then press “NEXT” to save the
setting and return you to the “Settings Menu”.
To disable this function enter a value of 0.0
o
C
6.4.3
Maintenance
The maintenance menu can be accessed via the settings menu.