OPERATING MANUAL FOR SELECT EDI 60
L300180 Iss:A
29
8
DISINFECTION PROCEDURE
Sanitisation of the unit should be carried out at least once every 6 months.
Switch the unit off by pressing the STOP button.
The unit will now have reverted to the POWER ON screen.
Turn off the water supply to the unit.
Leave the unit to stand for 60 seconds to relieve the pack water
pressure.
Pull away the front door cover to reveal the media cartridge packs.
Remove the left hand Pre-Treatment pack and discard.
o
To sanitise the unit insert a cleaning pack adaptor (Suez part
number L998549) into the left hand media pack location.
Turn on the water supply and press the menu button.
Select ‘Clean Routine’ from the menu and follow the on screen’
instructions.
When the clean routine has completed, isolate the water supply and
allow the unit to stand for 60 seconds to depressurise.
Remove the cleaning / descale pack and dispose of according to local
regulations.
Fit a new ‘Pre-Treatment pack and follow the ‘on screen’ instructions.
The unit will revert to the ‘Power On’ screen and be ready for operation