SPECIFICATIONS
13
F-20L Compressor
There are two electrical voltage variations:
Compressor P/N
Mains Voltage
280090D18N 208-230 VAC +/- 10%, V~, 60 Hz and
200 VAC +/- 10%, V~ 50 Hz
280090D27N 220 - 240 VAC +/- 10%, V~ 50 Hz
Electrical Characteristics
Service required: 2 poles, 3 wires (single phase plus protective ground.)
Mains Power Receptacle F-20L: Phoenix Heavycon Advance – Type HC-B10-TMS with screw
locks.
Power Consumption
Variation
Frequency
[Hz]
Voltage
Operation
[kW]
Compressor
RLA [A]
Compressor
LRA [A]
MCA
[A]
MOP
[A]
18N 60 220 2.6 12.0 56 15.0 30.0
50 200 2.4 12.0 15.0 30.0
27N 50 230 2.25 10.0 43-47 13.0 25.0
RLC = rated load current MCA = minimum circuit ampacity
LRA = locked rotor amps MOP = maximum overcurrent protection
Mains Power Receptacle: Phoenix Heavycon Advance – Type HC-B10-TMS with screw locks
Mains Power Supply Cable: 12 AWG, 3 conductor, 300V cordage with Phoenix connector and
NEMA L6-20P (2 pole, 3 wire, 20A, 250 V~) locking plug, P/N 280112C10
Circuit Protection:
Input: 20A, 250V circuit breaker
Control Circuit: (2) FU1 & FU2, 1.0A Slo Blo, 5 x 20 m fuses
Cold Head Valve Motor Circuit: (2) FU3 & FU4, 1.0A Slo Blo, 5 x 20 mm fuses
All fuses are located on front control panel