EasyManua.ls Logo

SuperFlow SF-1020I - Troubleshooting

SuperFlow SF-1020I
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
31
6.0 Maintenance
6.2 Troubleshooting
Before assuming anything is wrong with the owbench, verify that all of the electrical connections are
intact, and perform the operation ow test as detailed in "5.1.7 Performing a Flow Test" on page 19. If a
problem with the owbench measuring system arises, check the following:
Table 6-1. Troubleshooting
Problem Solution
The owbench will
not power on
Ensure the power cord is plugged in
Check the main breaker at the power distribution box
Check the 5-amp breaker in the power box behind the panel on the right
side of the owbench
The airow motors
shut o󰀨 after
running for a few
minutes
The motor is overheating—open the air bypass valve on the back of the
cabinet
One or more motors are defective and overheating—go to section "6.4.2
Blower Motor Replacement" on page 37.
The motor temperature sensor is defective or giving an incorrect
reading—go to section "6.4.3 Temperature Probe" on page 39.
Surging motor
Bi-stable phenomenon due to port in head
Line voltage varying
Motor is failing
Motor controller malfunction
The range selected
cannot be set
The range orice stepper motor is defective (see section "6.4.4 Orice
Stepper Motor Assembly" on page 39)
No Test Pressure
Reading
Check internal pressure transducer tubing. Ensure the hose has no kinks
or has not come o󰀨 of the transducer connection port.
The internal pressure transducer is defective or needs calibrating
No communication
between owbench
and the Modbus
TCP/IP network
Disconnected or defective cable
Modbus TCP/IP connection is not congured properly
Communication port malfunction on computer or control device
Internal PLC malfunction—contact SuperFlow Customer Service
No Test Pressure
Reading
Check internal pressure transducer tubing. Ensure the hose has no kinks
or has not come o󰀨 of the transducer connection port.
The internal pressure transducer is defective or needs calibrating
Flow readings are all
too high or too low
Perform the owbench ow tests according to the procedures in section
"6.3 Flowbench Flow Tests" on page 33.
Excessive leakage
Perform the no-ow test according to the procedure in section "6.3
Flowbench Flow Tests".

Table of Contents