6.4.7 Belt Length Calibration
Keep the belt weigher running at a constant speed. After the belt running time of a revolution being
measured by stopwatch, do this operation for correcting Belt Length.
New
LEN:
9.80m
Pulse: 980
7 Length Cal.
Old
LEN:
10.00m
ENT
[P102]: Belt Length [m]
Main Display Interface
MENU
+
: F2 Calibration
ENT
+
: 7 Length Cal.
ENT
M
Pulse: 35
Timer: 27.0s
7 Length Cal.
Length: 0.35m
0028.
0
[s]
R: 1.0-9999.9
Run-timer
Run-timer over or
ENT
MENU
: Exit
ENT
: Save
M