MAINTENANCE AND SUPPLIES REPLACEMENT
4-30 Sysmex KX-21 Operator’s Manual -- Revised October 1998
(4) Confirm that reagent in the transducer chamber has been drained.
RBC Transducer
WBC Transducer
Figure 4-7-10: Checking the WBC/RBC Transducer
(5) Remove the transducer chamber plug.
Plug
Figure 4-7-11: Removing the Plug
• A small volume of reagent occasionally remains. When reagent has
leaked, wipe it off immediately using wet cloth or the like. If left as it is, it
can cause current leakage or electric shocks.
(6) Apply CELLCLEAN on the brush provided, and dab by lightly prodding the brush
against the transducer aperture.
Brush
Aperture
Figure 4-7-12: Cleaning the Aperture
CAUTION