MAINTENANCE AND SUPPLIES REPLACEMENT
4-32 Sysmex KX-21 Operator’s Manual -- Revised October 1998
7.4 Replace Waste Tank
When the waste tank has become full, replace it by the following procedure:
When replacing the waste tank, always wear rubber gloves. After
replacing, wash hands with disinfectant. If your hands are
contaminated by blood, etc., infection of bacteria or the like can occur.
(1) Turn off the power of the main unit and wait approximately 30 seconds.
(2) Make ready an empty waste container and remove the cap.
(3) Pull the tube off the waste container.
(4) Insert the tube in the new waste container and fix it using tape or the like.
Tube
Waste
Tape
Figure 4-7-13: Waste Container
When using a used reagent container as the waste container, make
sure to clearly mark it that it is the waste container.
WARNING
CAUTION