MAINTENANCE AND SUPPLIES REPLACEMENT
Sysmex KX-21 Operator’s Manual -- Revised October 1998 4-33
8. SUPPLIES REPLACEMENT
8.1 Replenish Reagent
When reagent becomes insufficient during analysis, the instrument comes to a stop, while the
screen displays the error message. Press [HELP] key to change to the replenishment
operation.
Error
Message
(Analysis
Screen)
HELP Screen Reagent to
Replenish
Replenish
Diluent
*
Help
*
Replenish Diluent Container.
ERR CODE:XXXXX.X.X
Reaspirating Diluent[1]
1:Asp. Reag. 3:Cancel
CELLPACK
Replenish
Lyse
1:Asp.Reag.
ERR CODE:12490.0.0
*
Help
*
3:Cancel
Replenish Lyse Container.
Reaspirating Lyse[1]
STROMATOLYSER-WH
NOTE: Press [3] key on the HELP screen and select "3: Cancel" to return to
the Analysis screen. Although Analysis screen displays the Not
Ready status, the processing, etc. of stored data can be performed.