CALIBRATION
6-12 Sysmex KX-21 Operator’s Manual -- Revised October 1998
2. Updating Calibration Value
(1) When all calibration values are entered, press [SELECT] key.
The Calibration Value Change Confirmation message appears.
Change? Yes No
Figure 6-3-4: Calibration Value Change Confirmation Message
NOTE: • When there is no calibration value changed, the Calibration Value Change
Confirmation message is not displayed and the system returns to the
Analysis screen.
• When the compensation rate exceeds the following range, "Calibration
Error" is displayed and the Calibration Value Change Confirmation
message does not appear.
Compensation rate
=
Average of values gained by reference method
Average of values gained by this instrument
X 100
Compensation rate > 105%
Compensation rate < 95%
New compensation value > 120%
New compensation value < 80%
*
Manual Cal.
*
Calibration Error
Figure 6-3-5: Calibration Error Message
About calibration errors, refer to
Chapter 7: TROUBLESHOOTING.
(2) Using [ ] or [ ], move the cursor to select "YES" or "NO."
[YES]: Updates the calibration value and returns to the Analysis screen.
[NO]: Does not update the calibration value and returns to the Analysis screen.
(3) Press [ENTER] key to execute the selected process.
Sysmex KX-21 Operator’s Manual -- Revised December 1998