TROUBLESHOOTING
Sysmex KX-21 Operator’s Manual -- Revised October 1998 7-11
Error Message Meaning of Message
"Replenish Diluent"
ERR CODE:XXXXX.X.X
*
Help
*
1:Asp. Reag. 3:Cancel
Replenish Diluent Container.
Reaspirating Diluent
[1]
Diluent (CELLPACK) can not be aspirated
within the prescribed diluent aspiration time.
Cause 1) Lack of diluent (CELLPACK)
2) Diluent aspiration line error
3) Air leakage from vacuum line
4) Float switch malfunction
5) Faulty operation of solenoid valve or master valve
Action 1) Replenish Diluent (CELLPACK)
If there is no diluent, replenish with a new diluent container.
2) Check Tube
Check the diluent line tubing for breakage and the nipple for
looseness. If any abnormality is noted, reconnect or replace.
After replenishment or check, press [1] to return to the original
screen.
3) Check Vacuum Line
Check the vacuum line for breakage and loosened nipple or
tubing.
Analysis after error
occurrence
Not Ready until the error is resolved.
NOTE: Press [3] key to return to the Analysis screen. The status is not ready for
analysis, but the processing, etc. of stored data in Select Menu can be
performed.