TROUBLESHOOTING
7-12 Sysmex KX-21 Operator’s Manual -- Revised October 1998
Error Message Meaning of Message
"Replenish Lyse"
ERR CODE:XXXXX.X.X
*
Help
*
1:Asp. Reag.
3:Cancel
Replenish Lyse Container.
Reaspirating Lyse
[1]
No lyse remains.
Cause 1) Lack of lyse (STROMATOLYSER-WH)
2) Float switch malfunction
Action 1) Replenish Lyse (STROMATOLYSER-WH)
When no lyse is left, replenish with a new lyse container.
After replenishment or check, press [1] key to return to the original
screen.
Analysis after error
occurrence
Not Ready until the error is resolved.
NOTE: Press [3] key to return to the Analysis screen. The status is not ready for
analysis, but the processing, etc. of stored data in Select Menu can be
performed.