EasyManua.ls Logo

SYSMEX KX-21 - Page 201

SYSMEX KX-21
306 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TROUBLESHOOTING
Sysmex KX-21 Operator’s Manual -- Revised October 1998 7-27
3.9 Maintenance Errors
Error Message Meaning of Message
*
Scheduled Maint.
*
1:Exec.Clean
Clean the SRV.
Cycle No. after cleaning. 7567
Date last serviced 31/12/97
Reset the cycle counter for SRV,
and you can turn OFF the power.
[1]
[3]
Continue to start up. SRV cycle
counter will not be reset.
3:Cancel
It is time to clean the SRV.
Three months have passed from the previous
cleaning.
7500 samples have been analyzed from the
previous cleaning.
Cause
Action
Press [1] key. Following the instructions on the screen, turn OFF the
power and clean the sample rotor valve (SRV). (Refer to
Chapter 4:
Section 6.1: Clean Sample Rotor Valve (SRV)
).
If you perform cleaning later, press [3] key. This activates the usual
startup sequence and makes the system Ready.
The same message will appear the next time you start the unit; it will
be displayed at each startup until the cleaning is performed.
Analysis after error
occurrence
Ready for analysis
NOTE: When you clean SRV without pressing [1] key, execute "7: Maintenance,"
and then "3: Reset Counter" in the Select Menu.

Related product manuals